The CYCLONE 110 accelerates protons up to 65 MeV.

With light ions, the irradiation tests are made in air.

After passing through a scattering foil, air and the monitoring system, the maximum energy on the DUT is 62 MeV.

A set of 5 degradors enables to reach 17 different energies between 10 and 62 MeV.

Energy
[MeV]

LET
[MeV/(mg/cm2)]

Range
[mm]

10,0 (*)

3,47E-02

0,71

14,4

2,62E-02

1,34

20,5

1,99E-02

2,50

25,8

1,66E-02

3,76

30,1

1,47E-02

4,94

33,9

1,34E-02

6,10

37,3

1,21E-02

7,24

40,8

1,16E-02

8,49

43,8

1,10E-02

9,63

43,9

1,10E-02

9,67

46,8

1,04E-02

10,84

49,7

9,94E-03

12,06

52,3

9,56E-03

13,21

54,8

9,21E-03

14,35

57,2

8,92E-03

15,49

59,7

8,63E-03

16,71

62,0

8,39E-03

17,87

(*) For this energy, the FWHM (Full Duration at Half Maximum) is 5MeV

Beam parameters

The beam size diameter is 80mm with a homogeneity better than 10% within this area. 

The flux can be set from a few thousand of protons/(s.cm²) up to 2E8 protons/(s.cm²).